Ventilation, Mechanical

 Mechanical Ventilation

CLASSIFICATIONS OF POSITIVE-PRESSURE VENTILATORS

Modern ventilators are classified by the method of cycling from the inspiratory phase to the expiratory phase (ie, named after that parameter that signals the termination of the machine's positive-pressure inspiration cycle and changeover to the passive expiratory phase). The signal to terminate the machine's inspiratory activity can be a preset volume (volume-cycled ventilator), a preset pressure limit (pressure-cycled ventilator), or a preset time factor (time-cycled ventilator).

Volume-cycled ventilation is the most common form of ventilator cycling used in adult medicine because it provides a consistent breath-to-breath tidal volume. Termination of the delivered breath is achieved when a set volume has been delivered.

INDICATIONS FOR MECHANICAL VENTILATION

Many factors are considered in the decision to institute mechanical ventilation. Since no mode of mechanical ventilation can or will cure a disease process, the patient should have a correctable underlying problem that can be resolved while on mechanical ventilation. Mechanical ventilation should not be initiated without thoughtful consideration because intubation and positive-pressure ventilation could have potentially harmful effects.

Mechanical ventilation is indicated when the patient's spontaneous ventilation is not adequate to sustain life. In addition, it is indicated in critically ill patients to gain control of the patient's ventilation and as a prophylaxis for impending collapse of other physiologic functions. Physiologic indications are respiratory or mechanical insufficiency and gas exchange ineffectiveness. Some common indications for employing mechanical ventilation include the following:

Clinical judgment and the trend of these values should be used. A patient with an increasing illness severity should always alert a clinician to consider instituting mechanical ventilation.

INITIAL VENTILATOR SETTINGS

Mode of ventilation

After deciding to institute positive-pressure ventilation with a volume-cycled ventilator, the next decision is to select the most optimal mode of initial machine operation.

The initial mode of ventilation should be the assist control (A/C) mode. In this mode, a tidal volume and rate are preset, but whenever the patient makes an inspiratory effort, the ventilator senses the effort and delivers the preset tidal volume. In the A/C mode, the patient can initiate inspiration and control the breathing frequency; therefore, the respiratory pattern may have irregular intervals, depending on the timing of patient efforts. If the patient fails to initiate inspiration, the ventilator automatically goes into a back-up mode and delivers the preset rate and tidal volume, until it again senses an inspiratory effort. This back-up rate ensures minimum minute ventilation in the event of apnea. In the A/C mode, the work of breathing is reduced to only that amount of inspiration needed to trigger the machine's inspiratory cycle.

This mode is the closest to resting the respiratory muscles since the only work of breathing is the initial negative-pressure generation to trigger the machine. This mode has the advantage that the patient can increase minute ventilation by simply triggering additional breaths above the set back-up rate. In most cases, patients use their chemoreceptors and stretch receptors to achieve a minute ventilation that provides a reasonable pH. The respiratory center in the central nervous system receives input from the chemical receptors (arterial blood gas tensions) and neural pathways that sense the mechanical work of breathing (mechanoreceptors). The respiratory rate, respiratory pattern, and tidal volume are the result of input from these chemoreceptors and mechanical receptors. From this input, the respiratory center can regulate gas exchange with the minimum amount of work.

A second possible advantage is that cycling the ventilator into the inspiratory phase maintains normal ventilatory activity and, therefore, prevents atrophy of the respiratory muscles.

The potential disadvantages of the A/C mode include respiratory alkalosis. Because the ventilator delivers the set tidal volume on demand, a potential for alveolar hyperventilation exists, resulting in hypocapnia (observed most frequently in patients with end-stage liver disease, hyperventilatory stage of sepsis, and head trauma).

A second possible disadvantage is that every breath is a full, preset, positive-pressure breath that influences venous return to the right heart and, possibly, to global cardiac output. Nevertheless, the A/C mode is the best initial mode of ventilation and may be switched to another option if hypotension occurs or hypocarbia is evident on the first arterial blood gas analysis.

Tidal volume and rate

For a patient without preexisting lung disease, the tidal volume and rate traditionally are selected by the "12-12 rule". In the A/C mode, a tidal volume of 12 cc/kg of lean body weight is preset to be 12 times per minute.

In patients with chronic obstructive pulmonary disease (COPD), the tidal volume and rate are slightly reduced to the "10-10 rule" to avoid overinflation and hyperventilation. In the A/C mode, a tidal volume of 10 cc/kg of lean body weight is delivered 10 times per minute.

In acute respiratory distress syndrome (ARDS), new information suggests that the lungs function better and volutrauma is minimized with low tidal volumes of 6-8 cc/kg. In the A/C mode, tidal volumes are preset at 6 cc/kg of lean body weight at a rate of 8-10 breaths per minute. These lower volumes may lead to a slight degree of hypercarbia, which typically is recognized and accepted without correction, thus the term permissive hypercapnia.

Double checking the selected tidal volume

Once a tidal volume is selected, the peak airway pressure necessary to deliver a single breath should be recorded. As tidal volume increases, the pressure required to force that volume into the lung also increases. During mechanical ventilation, a persistent breath-to-breath peak pressure higher than 45 cm H2O is a risk factor for barotrauma. The tidal volume suggested by the above rules may need to be decreased in some patients to keep the peak airway pressure, if possible, below 45 cm H2O.

Sighs

Since a spontaneously breathing individual typically sighs 6-8 times per hour to avoid microatelectasis, giving periodic machine breaths of 1.5-2 times the preset tidal volume 6-8 times per hour once was recommended. Often the peak pressure needed to deliver such a volume would predispose to barotrauma.

Today, sighs usually are not used if the patient is receiving tidal volumes of 10-12 cc/kg or requires the use of positive end-expiratory pressure (PEEP). When using low tidal volumes, sighs are preset at 1.5-2 times the tidal volume and delivered 6-8 times per hour, provided the peak pressures are safe.

Initial FIO2

When initiating mechanical ventilation, the highest priority is to provide effective oxygenation. After intubation, 100% FIO2 always should be used until adequate oxygenation has been documented by postintubation and postmechanical ventilation arterial blood gases. A short period on FIO2 of 100% is not dangerous to the patient on mechanical ventilation and offers the clinician several advantages.

First, an FIO2 of 100% protects the patient against hypoxemia if unrecognized problems develop from the intubation procedure. With the PaO2 measured on an FIO2 of 100%, the clinician can easily calculate the next desired FIO2 and quickly estimate the shunt fraction.

The degree of shunt on 100% FIO2 can be estimated by a rough rule of thumb. The measured PaO2 is subtracted from 700 mm Hg. For each 100 mm Hg difference, a 5% shunt exists. A shunt of approximately 25% requires the use of PEEP.

Inadequate oxygenation despite the administration of 100% oxygen should prompt a search for complications of endotracheal intubation or positive-pressure breathing (mainstem intubation or pneumothorax). If such complications are not present, then PEEP is needed to treat intrapulmonary shunt pathology.

Since only a few disease processes can create an intrapulmonary shunt, the presence of a significant shunt narrows the patient's potential source of hypoxemia to alveolar collapse (major atelectasis) and alveolar filling with something other than gas (lobar pneumonia, ARDS, congestive heart failure [CHF], hemorrhage).

Positive end-expiratory pressure

PEEP is a mode of therapy used in conjunction with mechanical ventilation. At the end of exhalation (either mechanical or spontaneous), patient airway pressure is maintained above atmospheric pressure by exerting a pressure that opposes complete passive emptying of the lung. This pressure typically is achieved by maintaining a positive pressure flow at the end of exhalation. This pressure typically is measured in centimeters of H2O.

PEEP therapy can be effective when used in patients with a diffuse lung disease that results in an acute decrease in functional residual capacity (FRC), which is the volume of gas that remains in the lung at the end of a normal expiration. FRC primarily is determined by the elastic characteristics of the lung and chest wall. In many pulmonary diseases, FRC is reduced because of the collapse of the unstable alveoli. This reduction in lung volume decreases the surface area available for gas exchange and results in intrapulmonary shunting (ie, unoxygenated blood returning to the left side of the heart). If FRC is not restored, a high concentration of inspired oxygen may be required to maintain the arterial oxygen content of the blood in an acceptable range.

Applying continuous positive pressure at the end of exhalation (PEEP, continuous positive airway pressure [CPAP]) causes an increase in alveolar pressure and increases alveolar volume. This increase in lung volume increases the surface area by reopening or stabilizing collapsed or unstable alveoli. This “splinting” or “propping open” of the alveoli with positive pressure may provide a better matching of ventilation to perfusion, thereby reducing the shunt effect.

Once a true shunt is changed to a ventilation/perfusion (V/Q) mismatch, lower concentrations of oxygen can be used to maintain an adequate PaO2. PEEP therapy also has been reported to be effective in improving lung compliance. When FRC and lung compliance are decreased, more energy and volume are needed to inflate the lung. By applying PEEP, the lung volume at the end of exhalation is increased, which decreases the work of breathing because the lung is already partially inflated; therefore, less volume and energy are needed to inflate the lung.

In summary, when used to treat patients with a diffuse lung disease, PEEP should improve compliance, decrease dead space, and decrease the intrapulmonary shunt effect. The most significant benefit of PEEP is that the patient can maintain an adequate PaO2 at a lower, safer concentration of oxygen (<60%), thereby reducing the risk of oxygen toxicity.

Because PEEP is not a benign mode of therapy and can lead to serious consequences, the ventilator operator should have a definite indication to use it. Typically, the addition of external PEEP is justified when a PaO2 of 60 mm Hg cannot be achieved with a FIO2 of 60% or if the initial shunt fraction is greater than 25%. No recommendations exist for adding external PEEP during initial ventilator setup to satisfy misguided attempts to supply prophylactic PEEP or physiologic PEEP. Most clinicians use the lowest amount of positive pressure that provides an adequate PaO2 with a safe FIO2.

Summary of initial ventilator set up

ADJUSTING THE VENTILATOR AND AVOIDING COMPLICATIONS

After initiating ventilator support, the goal during the maintenance phase is to achieve adequate gas exchange without toxicity or complications, while allowing the primary cause of the respiratory failure to be treated or resolve. The first arterial blood gas typically is obtained 15-20 minutes after the patient is set up on the ventilator and dictates how the first changes in ventilation and oxygenation should be made.

Oxygenation decisions

The target PaO2 in respiratory failure is 60 mm Hg and/or 90% saturation of the hemoglobin. A PaO2 of 60 mm Hg usually is on the shoulder of the oxygen-hemoglobin saturation curve and typically is sufficient to ensure adequate oxygen delivery to the tissues.

When the clinician wishes to change the PaO2, 2 variables might be adjusted, the FIO2 and the amount of PEEP. When using supplemental FIO2, improvement in oxygenation cannot reliably be achieved with changes in rate or tidal volume.

With the target PaO2 identified, the FIO2 can be adjusted using the formula: New FIO2=(old FIO2 X desired PaO2)/measured PaO2.

PEEP principally is used to lower the risks of oxygen toxicity and is employed when a safe PaO2 cannot be achieved at 60% FIO2. Initiated at 5 cm, PEEP usually is increased in 3 cm H2O increments, while evaluating the effect on oxygenation every 15-20 minutes.

Ventilation decisions

In the early stages of mechanical ventilation, the target PaCO2 is based on the resultant pH. Errors in acid base rarely are clinically serious if mechanical ventilation is set up initially in the A/C mode because the pH is the primary stimulus to breathe. The PaCO2 is the value that results in a normalized pH. This is the correct PaCO2 because it results in a reasonable pH.

Occasionally, a patient may hyperventilate on the A/C mode and may develop a respiratory alkalosis (sepsis, liver disease, head trauma). Respiratory alkalosis with a pH greater than 7.5 may lead to cardiac arrhythmia when hypokalemia develops or may lead to seizure activity when cerebral blood flow is reduced by the vasoconstricting effects of hypocarbia. When this occurs, the clinician should first confirm that anxiety and pain are adequately controlled with pharmacotherapy. If hyperventilation persists, consider changing the mode of ventilation from assist control (where every breath is a positive-pressure breath of preset volume) to the synchronized intermittent mandatory ventilation (SIMV) or continuous positive airway pressure (CPAP) mode.

Synchronous intermittent mandatory ventilation

Intermittent mandatory ventilation (IMV) originally was developed for weaning patients from the ventilator. Later, this mode was found effective as a primary means of ventilating adult patients. Like the control mode, the ventilator delivers the mandatory breaths at a preset tidal volume and rate. When the patient breaths above the set rate on SIMV, the additional breaths are spontaneous breaths in tidal volumes that the patient is strong enough to pull. For the spontaneous breathing, the ventilator provides a gas source of oxygen of the same concentration, temperature, and humidity as is available to the patient for mandatory breathing. During the spontaneous phase of ventilation, the patient determines the respiratory rate and tidal volume. SIMV was developed to coordinate the mandatory breath delivered from the ventilator with the beginning of a patient's spontaneous inspirations when possible. Three types of breaths are possible on SIMV, as follows:

For the hyperventilating patient, the spontaneous breaths will be of smaller tidal volume than the preset machine breath, resulting in less alveolar wash out and less hypocarbia.

Continuous positive airway pressure

In extreme cases of hyperventilation, the clinician may default to the CPAP mode. With this mode, the ventilator relies on the patient to breathe spontaneously. The ventilator does not cycle or give any preset positive-pressure breaths; however, it can allow the clinician to maintain a higher than atmospheric end-exhalation pressure. When dialed in, this is termed CPAP and has the same physiologic characteristics as PEEP. When placed on the CPAP mode, the patient does all the work of breathing without the aid of a mechanical back-up rate from the ventilator. The ventilator circuit provides a gas source of oxygen at a prescribed concentration, temperature, and humidity. This mode of therapy primarily is used to wean the patient from mechanical ventilation and rarely is used as a stand-alone mode because it relies solely on the patient's respiratory drive. In a patient who is hyperventilating, a rapid spontaneous respiratory rate ensures that the spontaneous tidal volume will be small. Alveolar washout may beless and hypocarbia will resolve.

Paralysis

Occasionally, hyperventilation with significant respiratory alkalosis does not resolve on either of the above ventilator modes. In this situation, the patient is placed back on the A/C mode, with preset tidal volume and rate, and then paralyzed with a neuromuscular blocking agent. When performed, the clinician must realize that the patient's ventilation (PaCO2 and pH) is now completely controlled by the preselected rate and tidal volume. If adjustment in the PaCO2 and, thus, the pH is desired, the clinician must now limit attention to 1 of these 2 variables. In adult medicine, tidal volume usually is preserved and rate is adjusted using the formula: New rate=(old rate X measured PCO2)/desired PCO2.

Sedation decisions

Attention to the patient's anxiety and pain are imperative for successful mechanical ventilation. Benzodiazepines can be administered as bolus doses (as needed) but largely have been supplanted by the use of a continuous intravenous benzodiazepine drip. Pain usually is responsive to small intravenous boluses of narcotics (eg, morphine); however, an important caveat to remember is that if the patient is struggling against the ventilator, then the ventilator almost always can be set up in a better fashion. Checking and optimizing airway pressures and triggering sensitivity, inspiratory flow rate, and inspiratory-to-expiratory (I:E) ratio are wise before increasing sedation.

Peak and plateau pressures

In the ventilated patient, examining the peak and plateau pressures on the ventilator manometer often can identify the cause of acute respiratory distress. The peak pressure is the amount of pressure the machine requires to deliver the set tidal volume over the airway resistance and lung stiffness. To avoid barotrauma, this value should be kept below 40-45 cm H2O.

The plateau pressure is the amount of pressure required to keep the lung expanded after airflow is complete. The difference between peak and plateau pressure is an indication of the airway resistance. These values, when recorded daily and compared, often can help the clinician to understand why the ventilator is taking more pressure to deliver the same tidal volume. If both peak and plateau pressures increase, the lungs must be stiffer. If the peak pressure rises but the plateau does not, then the airway resistance has increased.

Inspiratory-to-expiratory ratio

In both spontaneous-assisted and ventilator-assisted breathing, exhalation is a passive event, depending on the elastic properties of the thorax. In a healthy person, the expiratory time is at least twice the inspiratory time, resulting in an inspiratory-to-expiratory (I:E) ratio of 1:2. In conventional ventilation, patient comfort is important to allow an I:E of at least 1:2. When ventilating a patient with chronic obstructive lung disease, an I:E ratio of 1:3 or 1:4 may be necessary.

If the I:E ratio is not preserved, the inspired gas is not allowed enough time to exit the lung before the next breath is given. Subsequently, breaths become stacked and intrathoracic pressure increases, resulting in positive pressure at end-expiration (ie, intrinsic PEEP).

With mechanical ventilation, the inspiratory time can be adjusted by changing the peak inspiratory flow rate. If the clinician wishes to change the I:E ratio and allow more time for exhalation, the peak inspiratory flow rate can be increased or the overall tidal volume can be decreased. The largest tidal volume possible without excessive peak inspiratory pressure (PIP), the slowest respiratory rate consistent with effective CO2 elimination, and inspiratory flow rates fast enough to decrease the I:E ratio all tend to reduce intrinsic PEEP.

Adjusting external positive end-expiratory pressure

A PEEP level of 10 cm H2O rarely causes hemodynamic problems in the absence of intravascular volume depletion. The cardiodepressant effects of PEEP often can be minimized by judicious intravascular volume support or cardiac inotropic support. While peak pressure is related to the development of barotrauma, arterial hypotension is related to the mean airway pressure, causing a decrease in either venous return to the heart or right ventricular function.

A PEEP level greater than 10 cm H2O generally is an indication to follow cardiac output (CO) with a Swan-Ganz catheter. The left atrial filling pressure (LAP) should be corrected for a PEEP greater than 10 cm: LAP=(pulmonary capillary wedge pressure [PCWP]-[PEEP/2]).

Withdrawal of PEEP from a patient should not be performed until the patient has satisfactory oxygenation on 40%. A formal PEEP wean, then, is performed by reducing the PEEP in 3- to 5-cm decrements while watching the hemoglobin-oxygen saturations. An unacceptable drop in the hemoglobin-oxygen saturation should prompt the clinician to immediately reinstitute the last PEEP level that provided good hemoglobin-oxygen saturation.

NEWER MODES OF MECHANICAL VENTILATION

A better understanding of ventilator-induced lung injury recognizes that the alveolar epithelium is at risk of both barotrauma and volutrauma. Barotrauma refers to rupture of an alveolus with subsequent air entry into the pleural space (pneumothorax) and/or air tracking back along the vascular bundle to the mediastinum (pneumomediastinum). Volutrauma refers to the lung injury that can occur when alveoli are overdistended and produce proinflammatory cytokines that augment or perpetuate the initial lung injury that led to mechanical ventilation.

With computer feedback, many modern ventilators allow the operator to make fine adjustments in tidal volume, airway pressures, and timing of the respiratory cycle in attempts to limit ventilator-induced lung injury. These methods of mechanical ventilation often are based on attractive physiologic hypotheses and are interesting to perform. While each method has its proponents, objective evidence fails to show that any one of the alternative ventilation methods is more successful than conventional mechanical ventilation with proper attention to tidal volume.

Currently, most clinicians employ alternative ventilation methods only when conventional mechanical ventilation strategy fails, which is rare.

Bibliography:

Home page, Drug information's ,Regional anesthesia ,Index for  diseases ,Index for  diseases, Perioperative medicine, Search engine, Accreditation, Learn Arabic Warfare, Anesthesia LINKS, Orthopedic LINKS, Midline

Constructed by Dr N.A. Nematallah Consultant in perioperative medicine and intensive therapy, Al Razi Orthopedic Hospital , State of Kuwait, email : razianesth@freeservers.com